PDV Portable Digital Vibrometer - Technical Data

General Specifications

PDV-100 General Specifications

Type of measurement

Velocity

Signal processing

Digital

Frequency range

0 - 22 kHz

Measurement ranges

3

Peak velocity (mm/s)

± 20

±100

±500

Scaling factor (mm/s/V)

5

25

125

Velocity resolution (1) (µm/s rms)

< 0.05

< 0.1

< 0.3

Working distance (2)

0.2 to 30 m

Laser safety

Eye safe class II visible HeNe laser

PDV-100 Output Signals

Analog velocity output

      Output voltage range

± 4 V, 24 bit DAC

      Frequency range

0.5 Hz - 22 kHz

      Dynamic range (3)

> 90 dB

     Calibration accuracy

±1 % (20 Hz - 22 kHz)

Digital velocity output

      Electrical S/P-DIF (4) Interface

24 bit, 48 kSa/s

      Frequency range

0 - 22 kHz

      Calibration accuracy

± 0.2 % (0.05 Hz - 22 kHz)

Output filter

Digital low pass filter (FIR type)

1, 5, 22 kHz (-0.1dB), roll-off 120 dB/dec

Analog high pass filter

100 Hz (-3dB), roll-off 60 dB/dec

PDV-100 Housing and Power

Dimensions

300 (L) x 63 (W) x 129 (H) mm

Weight

2.6 kg

Protection rating

IP-64 (dust and spray water protected)

Operating temperature

+5 to +40 °C

Relative humidity

max. 80% non-condensing

Display

Illuminated 3 line LCD

Power

11...14.5 V DC, max 15 W

Battery Kit PDV-BS

Rechargeable Li-Ion battery for min. 4 hours operation time

Remarks:

(1) w/ SNR 0dB 10 Hz RBW on 3M Scotchliteâ Tape

 (2) The maximum stand-off distance depends on the surface properties of the object.

(3)  Definied as spurious free dynamic range (SFDR)

(4) S/P-DIF: Sony/ Philips Digital Audio Inter Face

  

    
Compliance with Standards

 

Electrical safety

EN 61010 (IEC 1010)

EMC Emission

EN 50081-1 (FCC Class B)

EMC Immunity

EN 50082-1, EN/IEC 61000-6-2

Laser safety

EN/IEC 60825-1 (CFR 1040.10, 1040.11)

CE

Mark (EMC, laser safety, LVD)