VDD Digital Vibrometers - Technical Data
Front-End Specifications
|
VDD-E-600 Stand-Alone Digital Front-End |
|
Power |
100 VAC ... 240 VAC, max. 75 W |
|
Dimensions [W x L x H] |
235 mm x 320 mm x 150 mm (9.3 in x 12.6 in x 5.9 in) |
|
Weight |
5.7 kg (12.6 lbs) |
|
Signal inputs REF1/REF2 |
±200 mV ... ±10 V, into 1 MW |
|
Input coupling |
AC or DC, adjustable by software |
|
Inputs TRIG, GATE, AUX |
TTL/CMOS |
|
Output SIGNAL |
Internal generator output, max. ±10 V |
|
Outputs SYNC, AUX |
TTL/CMOS |
|
DD-600 Signal Converter Board for OFV-5000 Vibrometer Controller |
|
Decoder type |
Auxiliary decoder (I&Q converter) |
|
Output signal |
Baseband Doppler signal in I&Q format for external VDD decoding |
|
VDD-Z-011 Junction Box |
|
Signal input REF |
±200 mV ... ±10 V, into 1 MW |
|
Input coupling |
AC or DC, adjustable by software |
|
Inputs/outputs |
as specified for VDD-E-600 |
Performance Specifications
VDD Performance Specifications |
|
Frequency range |
0 Hz … 2 MHz |
|
Max. no. of FFT lines |
12,800 |
|
Maximum no. of samples |
- Time mode: 64 MSa
- FFT mode: 32,768
|
|
Max. velocity for frequency, f, and sample time, t |
Frequency |
f £ 51.2 kHz |
f > 51.2 kHz |
|
Time t < 1 s |
810 mm/s |
810 mm/s |
|
Time t = 1 s ... 2 s |
810 mm/s |
405 mm/s |
|
Time t > 2 s |
405 mm/s |
405 mm/s |
|
Velocity resolution1), depending on the vibration frequency f |
f = 2 kHz |
f = 20 kHz |
f = 200 kHz |
f = 2 MHz |
|
0.005 µm s-1/ÖHz |
0.05 µm s-1/ÖHz |
0.5 µm s-1/ÖHz |
5 µm s-1/ÖHz |
|
Max. displacement |
unlimited |
|
Displacement resolution1) |
< 0.4 pm/ÖHz |
|
Maximum displacement error |
- ±(1 % of RMS reading) for 1 nm … 100 nm
- ±(0.2 % of RMS reading + 1 nm) for ³ 100 nm
|
|
Total harmonic distortions |
< 0.1 % up to 50 kHz (sinusoidal vibration) |
|
Amplitude frequency response |
- Displacement: typ. ±0.1 %
- Velocity: typ. ±0.2 %
- Acceleration: typ. ±0.2 %
|
|
Spurious signals |
<20 pm |
1) Resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0dB in a 1 Hz resolution bandwidth (RBW), measured on 3M Scotchlite Tape™ (reflective film). The velocity resolution values refer to the settings for a maximum velocity of 810 mm/s.
Data Acquisition System
|
VDD Data Acquisition System |
|
Acquisition board |
4 channels (2 for I/Q signal, 2 for reference); 5 MSa/s, 12 bits |
|
Computer (optional) |
PC-D Desktop PC or PC-I Industrial PC (from Polytec)
System requirements: AMD Athlon™ 3000 XP+ or Intel Pentium 4; 3.0 GHz or higher; 512 MByte RAM; Windows® XP Professional or Windows® 2000 operating system |
|
Signal generator (optional) |
1 channel, 512 kHz, arbitrary waveforms |
|
External signal generator |
HP-33250A (option from Polytec) or HP-33120A (discontinued model), connected to GPIB via USB adapter |
VibSoft-VDD Software
|
VibSoft-VDD Software |
|
Vib-S-VDD
|
Digital data acquisition and processing, time & frequency domain displays, time & frequency domain averaging, digital filter program, time domain integration & differentiation. Input channels for vibrometer & reference (VDD-E-600: 2 references) |
|
VIB-S-GateIn |
TTL Gate input software |
|
VIB-S-FFT128 |
Supports 12,800 FFT lines |
|
VIB-S-VBEng
|
Creates Visual Basic® Scripts for automation of measurements and data presentation. Includes User Defined Data Sets (UDDS) for adding external or post processed data |
|
VIB-S-ExpUFF |
Converts data to Universal File Format (UFF) |
|
VibSoft-VDD Options |
|
VIB-S-SIG-M |
Internal arbitrary waveform generator |
|
VIB-S-FFTEXT |
Allows calculation of extended amount of FFT lines up to 819,200 lines* |
|
VIB-S-ExtSig |
Controls external arbitrary waveform generator via optional adapter |
|
VIB-S-ExpME |
Binary data interface for data exchange with ME’scope software |
|
VIB-S-SigPro |
Polytec Signal Processor, the user interface to the math library included in the VibSoft Software |
|
VIB-S-VIDEO |
Video option for VibSoft and VDD Software |
*)Possible number of FFT lines depends on the sampling rate of the acquisition system.
Compliance with Standards
|
Compliance with Standards |
|
Electrical safety |
IEC/EN 61010 |
|
EMC |
IEC/EN 61326
Emission: FCC Class B, IEC/EN 61000-3-2 and 61000-3-3
Immunity: IEC/EN 61000-4-2 to 61000-4-6 and IEC/EN 61000-4-11 |
|
Laser safety |
IEC/EN 60825-1 (CFR 1040.10, CFR 1040.11) |
|
 |
|