HSV-2000 High Speed Vibrometer - Technical Data

Performance Specifications

 

 

Velocity

Displacement

Frequency range

0… 50 kHz

0... 250 kHz (small signal)

Output swing

± 6 V

± 8 V

Scaling factor

5 m/s/V

0.32; 0.64; 1.28; 2.56; 5.12; 10.24 [mm/V]

Full scale

± 30 m/s

5.12;10.24; 20.48; 40.96; 81.92; 163.84* [mm p-p]

Resolution

< 15 µm/s**

0.32; 0.64; 1.28; 2.56; 5.12; 10.24 [µm] ***

Max. acceleration

1,000,000 g

45,000,000 g****

*  VibSoft-FC extends the maximum range to virtually unlimited

** The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0dB in a 10Hz spectral bandwidth (BW), measured at 3M Scotchlite Tape® (reflective film).

*** The resolution is defined as 1 increment of the fringe counter output, which is equivalent to a 1mV output voltage step.

**** At 10.24 mm/V, with the tracking filter switched off.

    
Optics

 

Laser output power

< 1 mW (eye safe class 2)

Stand-off distance

  • Variable focus: 0.11 m … 10 m
        (depending upon surface scattering properties)
  • Fixed focus: 520 mm

    
Housing and Operational Conditions

 

 

System component

HSV-2001/2002 Controller

HSV-800
Laser Unit*

HSV-700
Sensor Head

Dimensions
[L x W x H]

450 mm x 355 mm x 135 mm

(17.7 in x 14.0 in x 5.3 in)

(19’’ Rack mount)

340 mm x 130 mm x 115 mm

(13.4 in x 5.1 in x 4.5 in)

174 mm x 48 mm x 39 mm

(6.9 in x 1.9 in x 1.2 in)

(Fixed focus version)

Weight

8 kg (17.6 lb)

6 kg (13.2 lb)

0.5 kg (1.1 lb)

Protection class

I (protective ground)

IP 64

IP 64

Operating temperature

+5 °C ... +40 °C (41 °F ... 104 °F)

Storage temperature

-10°C ... +65 °C (14 °F ... 149 °F)

Relative humidity

max. 80%, non-condensing

Connection cables

Sensor Head to Laser Unit: standard 3 m (9.8 ft)

Laser Unit to Controller:        standard 5 m (16.4 ft)
                                        optional up to 10 m (32.8 ft)

Mains voltage

100 VAC ... 240 VAC ±10%, 50/60 Hz       

Power consumption

max. 100 VA

* For differential operation with using the HSV-2002, two laser units are required.

    
Compliance with Standards

 

 

Electrical safety

IEC/EN 61010-1

EMC

IEC/EN 61326

Emission: FCC Class A, IEC/EN 61000-3-2 and 61000-3-3

Immunity: IEC/EN 61000-4-2 to 61000-4-6 and IEC/EN 61000-4-11

Laser safety

IEC/EN 60825-1 (CFR 1040.10, CFR 1040.11)