Processing scheme |
Analog frequency demodulation |
|
Measurement ranges |
DC velocity:-40 m/s ... +50 m/s, slope: (5 m/s per V)
AC velocity in 4 ranges:
0.025 m s-1/V, 0.1 m s-1/V, 0.5 m s-1/V, 5 m s-1/V |
|
Resolution (AC, 0.025 m×s1/V) |
<25 µm/s in a 10 Hz spectral bandwidth |
|
AC-frequency response
(Filters OFF) |
Range: 0.025 m s-1/V: 0.5 Hz ... 50 kHz : ± 1 dB
Range:0.1 m s-1/V: 0.5 Hz ... 50 kHz : ± 0.5 dB
Range:0.5 m s-1/V: 0.5 Hz ... 50 kHz : ± 0.5 dB
Range:5 m s-1/V: 0 Hz ... 50 kHz : ± 0.5 dB |
|
Maximum acceleration |
500,000 m/s2 (Tracking Filter FAST) |
|
Signal output velocity DC |
Voltage swing: +10/-8 Volt,
Output impedance: 100 W
Load impedance: min 10 kW (-1% additional calibration error)
Calibration accuracy: ± 2% of reading |
|
Signal output velocity AC |
Voltage swing: ± 10 Volt
Output impedance: 50 W
Load impedance: min 10 kW (-0.5% additional calibration error)
Calibration accuracy: ± 2.5% of RMS reading |
|
Filters |
High pass 1 Hz ... 255 Hz (28settings), 4th order
Butterworth
Low pass 200 Hz ... 51 kHz (28 settings), 4th order Butterworth |
|
Data display |
Illuminated LCD for display of DC velocity and
instrument settings |
|
Interfaces |
RS-232 and IEEE-488
Input Monitor: test output for the sensor Doppler signal (40 MHz)
Output Monitor: test output for the conditioned Doppler
signal (40 MHz) |
|
Line Power |
100/115/230 volt, 50/60 Hz |
|
Power consumption |
Max. 150 VA |
|
Dimensions |
450 mm x 360 mm x 145 mm (W x L x H) |
|
Mass |
11 kg |
|
Operating temperature |
+5°C... 40°C |