OFV-2500 Series Vibrometer Controllers - Tech. Data
General Specifications
|
| Controller type |
OFV-2500 |
OFV-2510 |
OFV-2502 |
| Dimensions (L x W x H) |
235 mm x 320 mm x 150 mm (½ 19”, 42 HP/3 U) |
235 mm x 320 mm x 150 mm (½ 19”, 42 HP/3 U) |
450 mm x 360 mm x 135 mm (19“, 84 HP/3 U) |
| Analog signal outputs |
-Velocity -Displacement (integrator option) -Signal strength (0…+5 V) |
- Displacement - Signal strength (0…+5 V) |
- Velocity A/B - Signal strength A/B (DC 0…+5 V) |
| Digital interfaces |
- RS-232 |
- RS-232 - TTL interface to external fringe counter - TTL trigger input |
- USB 1.1 interface to control the vibrometer |
| Weight |
5.6 kg |
5.5 kg |
9 kg |
| Power |
100 VAC ... 240 VAC ±10 %, 50/60 Hz, max. 75 W |
| Ambient temperature |
+5 °C … +40 °C (41 °F … 104 °F) |
| Storage temperature |
-10 °C … +65 °C (14 °F … 149 °F) |
| Relative humidity |
max. 80 %, non-condensing |
| Compatibility |
OFV-505/503, OFV-551/552, OFV-534 Sensor Heads 1) | 1) When operating the OFV-534 Sensor Head with the OFV-2510 and OFV-2502 Controllers, the signal level display on the sensor is not available.
OFV-2500-2 Performance Specifications
| |
| Velocity Decoder |
| Range |
Full scale (peak) |
Lower frequency limit |
Upper frequency limit |
Resolution (typical) 1) |
Acceleration |
Linearity error |
| mm s-1/V |
m/s |
Hz |
kHz |
µm s-1/√Hz |
g |
± % |
| 10 |
0.1 |
0.5 |
250 |
0.2 |
16,000 |
1 |
| 100 |
1.0 |
0.5 |
3,000 |
0.5 |
1,900,000 |
1 |
| 1,000 |
10(7) 2) |
0.5 |
3,200 |
2.5 |
15,300,000 |
1 |
| Displacement Decoder (Optional Integrator) |
| µm/V |
µm |
Hz |
kHz |
nm |
g |
± % |
| 0.01 |
0.1 |
1,000 |
250 |
depends on velocity range setting |
depends on velocity range setting |
2.5 |
| 0.1 |
1 |
1,000 |
250 |
2.5 |
| 1 |
10 |
1,000 |
250 |
2.5 |
| 10 |
100 |
100 |
250 |
2.5 |
| 100 |
1,000 |
10 |
20 |
2.5 |
| 1,000 |
10,000 |
10 |
20 |
2.5 |
| 10,000 |
100,000 |
10 |
20 |
2.5 | 1) The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0dB with 1 Hz spectral resolution, measured on 3M Scotchlite Tape® (reflective film). The typical value refers to the center of the operating frequency range. 2) 10 m/s @ 300 kHz; 7m/s @ 3200 kHz
OFV-2500-3 Performance Specifications
| |
| Velocity Decoder |
| Range |
Full scale (peak) |
Lower frequency limit |
Upper frequency limit |
Resolution (typical) 1) |
Max. acceleration |
Linearity error |
| mm s-1/V |
mm/s |
Hz |
kHz |
µm s-1/√Hz |
g |
± % |
| 2 |
20.0 |
0 |
100 |
0.02 |
1,280 |
< 0.1 |
| 5 |
50.0 |
0 |
100 |
0.02 |
3,200 |
< 0.1 |
| 10 |
100 |
0 |
350 |
0.05 |
22,000 |
< 0.1 |
| 20 |
200 |
0 |
350 |
0.06 |
44,000 |
< 0.1 |
| 50 |
500 |
0 |
350 |
0.06 |
110,000 |
< 0.1 | 1) The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0dB with 1 Hz spectral resolution. measured on 3M Scotchlite Tape® (reflective film). The typical value refers to the center of the operating frequency range.
OFV-2510 Performance Specifications
|
| Displacement Decoder |
| Range |
Full scale (peak to peak) 1) |
Lower Amplitude limit |
Upper frequency limit 2) |
Resolution (typical) 3) |
Max. acceleration |
Linearity error 4) |
| µm/V |
µm |
µm |
Hz |
µm/count |
g |
± % |
| 80 |
1,280 |
0.4 |
800 |
0.08 |
14,500 |
1 |
| 160 |
2,560 |
0.8 |
400 |
0.16 |
14,500 |
1 |
| 320 |
5,120 |
1.5 |
200 |
0.32 |
14,500 |
1 |
| 640 |
10,240 |
3.0 |
100 |
0.64 |
14,500 |
1 |
| 1280 |
20,480 |
6.0 |
50 |
1.28 |
14,500 |
1 |
| 2,560 |
40,960 |
12.0 |
25 |
2.56 |
14,500 |
1 |
| 5,120 |
81,920 |
25.0 |
12 |
5.12 |
14,500 |
1 |
1) Full scale values correspond to ±8 V maximum output voltage. 2) Maximum frequency at 50% of full output amplitude. The decoder frequency limit which is valid for vibration amplitudes up to 1 µm is 250 kHz. 3) The resolution is defined as 1 increment of the fringe counter output, which corresponds to an output voltage step of 1 mV. 4) ±1 increment
OFV-2502 Performance Specifications
|
| Velocity Decoder |
| Range |
Full scale (peak) |
Lower frequency limit |
Upper frequency limit |
Resolution (typical) 1) |
Max. acceleration |
Linearity error |
| mm s-1/V |
M/s |
Hz |
kHz |
µm s-1√Hz |
g |
± % |
| 5 |
0.05 |
0.5 |
250 |
0.1 |
8,000 |
1 % |
| 25 |
0.25 |
0.5 |
1,500 |
0.3 |
240,000 |
1.5 % |
| 125 |
1.25 |
0.5 |
1,500 |
0.4 |
1,200,000 |
1 % |
1) The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0dB with 1 Hz spectral resolution, measured on 3M Scotchlite Tape® (reflective film). The typical value refers to the center of the operating frequency range.
Compliance with Standards
| |
| Laser safety |
IEC/EN 60825-1 (CFR 1040.10, CFR 1040.11) |
| Electrical safety |
IEC/EN 61010-1 |
| EMC |
IEC/EN 61326 |
|
 |
OFV-2500 Vibrometer Controller
OFV-2510 Vibrometer Controller
OFV-2502 Vibrometer Controller
|