IVS-300 Industrial Vibration Sensor - Technical Data
General Specifications
|
|
|
Ambient temperature |
+5 °C … +40 °C |
|
Storage temperature |
-5 °C … +60 °C |
|
Relative humidity |
20 % – 80 %, non-condensing |
|
Housing protection |
IP 64 standard |
|
Dimensions |
302 mm x 114 mm x 55 mm |
|
Weight |
2.6 kg |
|
Power |
11 V – 14.5 V DC, max. 15 W |
|
Decoder type |
DSP velocity decoder, 3 measurement ranges |
|
Max. velocity |
0.5 m/s peak |
|
Velocity ranges |
20 mm s-1 / 100 mm s-1 / 500 mm s-1 (full scale) |
|
Velocity resolution |
< 0.05 µm s-1 / ÖHz |
|
Frequency range |
0 – 22 kHz |
|
Filters
|
Digital low pass filter 1kHz / 5 kHz / 22 kHz (-1 dB), roll-off >120 dB/dec (analog and digital output)
Analog high pass filter 100 Hz (-3 dB) /off, roll-off about 60 dB/dec (analog output only) |
|
Output analog |
± 4 V, 24-bit DAC, ranges: 5 / 25 / 125 (in mm s-1/V) |
|
Output digital |
S/P-DIF (SONY/Philips Digital Audio Interface) 24 bit, 48 kSa/s |
|
Connectors |
Industrial connector for voltage supply, optical signal level and velocity output
Connector for IVS-310 signal level indicator and RS-232 cable
Triax connector for S/PDIF digital output |
|
Optical system |
Fixed focus lens (ff): 240 mm stand-off distance
Variable focus lens (vf): 90 mm to 3 m stand-off distance |
|
Laser safety |
< 1 mW output power, laser safety class II, visible 632.8 nm laser |
| |
|
 |
Compliance with Standards
|
|
|
Electrical safety |
EN 61010 (IEC 1010) |
|
EMC Emission |
EN 50081-1 (FCC Class B) |
|
EMC Immunity |
EN 50082-1, EN/IEC 61000-6-2 |
|
Laser safety |
EN/IEC 60825-1 (CFR 1040.10, 1040.11) |
|
CE |
Mark (EMC, laser safety, LVD) |
|
 |
|
|