VDD Digital Vibrometers - Technical Data

Front-End Specifications

VDD-E-600 Stand-Alone Digital Front-End

Power

100 VAC ... 240 VAC, max. 75 W

Dimensions [W x L x H]

235 mm x 320 mm x 150 mm (9.3 in x 12.6 in x 5.9 in)

Weight

5.7 kg (12.6 lbs)

Signal inputs REF1/REF2

±200 mV ... ±10 V, into 1 MW

Input coupling

AC or DC, adjustable by software

Inputs TRIG, GATE, AUX

TTL/CMOS

Output SIGNAL

Internal generator output, max. ±10 V

Outputs SYNC, AUX

TTL/CMOS

DD-600 Signal Converter Board for OFV-5000 Vibrometer Controller

Decoder type

Auxiliary decoder (I&Q converter)

Output signal

Baseband Doppler signal in I&Q format for external VDD decoding

VDD-Z-011 Junction Box

Signal input REF

±200 mV ... ±10 V, into 1 MW

Input coupling

AC or DC, adjustable by software

Inputs/outputs 

as specified for VDD-E-600

    
Performance Specifications

 


VDD Performance Specifications

Frequency range

0 Hz … 2 MHz

Max. no. of FFT lines

12,800

Maximum no. of samples

  • Time mode: 64 MSa
  • FFT mode: 32,768

Max. velocity for frequency, f, and sample time, t

Frequency

£ 51.2 kHz

f > 51.2 kHz

Time t < 1 s

810 mm/s

810 mm/s

Time t = 1 s ... 2 s

810 mm/s

405 mm/s

Time t > 2 s

405 mm/s

405 mm/s

Velocity resolution1), depending on the vibration frequency f

f = 2 kHz

f = 20 kHz

f = 200 kHz

f = 2 MHz

0.005 µm s-1/ÖHz

0.05 µm s-1/ÖHz

0.5 µm s-1/ÖHz

5 µm s-1/ÖHz

Max. displacement

 unlimited

Displacement resolution1)

< 0.4 pm µm s-1/ÖHz

Maximum displacement error

  • ±(1 % of RMS reading) for 1 nm … 100 nm
  • ±(0.2 % of RMS reading + 1 nm) for ³ 100 nm

Total harmonic distortions

< 0.1 % up to 50 kHz (sinusoidal vibration)

Amplitude frequency response

  • Displacement: typ. ±0.1 %   
  • Velocity: typ. ±0.2 %
  • Acceleration: typ. ±0.2 %

Spurious signals

<20 pm


1)
Resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0dB in a 1 Hz resolution bandwidth (RBW), measured on 3M Scotchlite Tape™ (reflective film). The velocity resolution values refer to the settings for a maximum velocity of 810 mm/s.

    
Data Acquisition System

VDD Data Acquisition System

Acquisition board

4 channels (2 for I/Q signal, 2 for reference); 5 MSa/s, 12 bits

Computer (optional)

PC-D Desktop PC or PC-I Industrial PC (from Polytec)

System requirements: AMD Athlon™ 3000 XP+ or Intel Pentium 4; 3.0 GHz or higher; 512 MByte RAM; Windows® XP Professional or Windows® 2000 operating system

Signal generator (optional)

1 channel, 512 kHz, arbitrary waveforms

External signal generator

HP-33250A (option from Polytec) or HP-33120A (discontinued model),
connected to GPIB via USB adapter

    
VibSoft-VDD Software

VibSoft-VDD Software

Vib-S-VDD

Digital data acquisition and processing, time & frequency domain displays, time & frequency domain averaging, digital filter program, time domain integration & differentiation. Input channels for vibrometer & reference (VDD-E-600: 2 references)

VIB-S-GateIn

TTL Gate input software

VIB-S-FFT128

Supports 12,800 FFT lines

VIB-S-VBEng

Creates Visual Basic® Scripts for automation of measurements and data presentation. Includes User Defined Data Sets (UDDS) for adding external or post processed data

VIB-S-ExpUFF

Converts data to Universal File Format (UFF)

VibSoft-VDD Options

VIB-S-SIG-M

Internal arbitrary waveform generator

VIB-S-FFTEXT

Allows calculation of extended amount of FFT lines up to 819,200 lines*

VIB-S-ExtSig

Controls external arbitrary waveform generator via optional adapter

VIB-S-ExpME

Binary data interface for data exchange with ME’scope software

VIB-S-SigPro

Polytec Signal Processor, the user interface to the math library included in the VibSoft Software

VIB-S-VIDEO

Video option for VibSoft and VDD Software

*)Possible number of FFT lines depends on the sampling rate of the acquisition system.

    
Compliance with Standards

Compliance with Standards

Electrical safety

IEC/EN 61010

EMC

IEC/EN 61326

Emission: FCC Class B, IEC/EN 61000-3-2 and 61000-3-3

Immunity: IEC/EN 61000-4-2 to 61000-4-6 and IEC/EN 61000-4-11

Laser safety

IEC/EN 60825-1 (CFR 1040.10, CFR 1040.11)